Hospitech has offered uncompromised Ethylene Oxide sterilization service and quality since 1998. Our company has many years of industrial experience in the medical device, furniture and glove industry. We adhere to strict compliance of the ISO11135 & EN550 standards. Our key objective is to guide our customers the entire sterilization process to ensure hassle-free, prompt and accurate results.

The most crucial parameters in Ethylene Oxide Sterilization are the temperature, pressure and humidity. Our sterilization process employs 9 controlled steps as follows:

Air Leak Test
Chamber Leak Test
Deep Vacuum
Humidification
Gas Admission
Sterilization
Pulse Post Vacuum
Aeration
Equalization
 
Quick Facts of Sterilization in Hospitech:
   
Two (2) Chambers to Cater Sterilization Demand, with a capacity of approximately 16cbm and 12cbm respectively.
Accommodates 8 pallets for each Chamber.
Pallet Size: 800X1200X200mm.
Gas Composition: 90% Ethylene Oxide, 10% Carbon Dioxide.
Capable of Handling All Product Volumes – from Single Cases to Containers.
Provides Validation, from the planning stages to complete validation report.
Customized EtO Cycles, based on Customer’s Requirements.

 
 